• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL ADVISA DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL ADVISA DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number A5DR01
Device Problem Premature Discharge of Battery (1057)
Patient Problem Exit Block (2628)
Event Date 08/10/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced intermittent heart block.It was reported that a possible fracture and high and unstable thresholds were noted on the right ventricular (rv) lead.It was reported that the lead fracture may have hastened battery depletion on the implantable pulse generator (ipg).The lead was capped and replaced and the ipg was explanted and replaced.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory showed the battery is approaching the indicator signifying the time for device replacement.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISA DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
MDR Report Key10481776
MDR Text Key205262868
Report Number9614453-2020-02704
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/14/2012
Device Model NumberA5DR01
Device Catalogue NumberA5DR01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/11/2020
Initial Date FDA Received09/02/2020
Supplement Dates Manufacturer Received09/03/2020
Supplement Dates FDA Received09/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
5076-58 LEAD, 5076 LEAD; 5076-58 LEAD, 5076 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age92 YR
Patient Weight92
-
-