H10: h2, h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.No product identification information was provided and thus a manufacturing record review could not be conducted.A review of the instructions for use found as with any surgical instrument, care should be taken to ensure that excessive force is not placed on the device.Excessive force can result in failure.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.Factors that could have contributed to the reported event include (1) excessive force.(2) tissue thickness.(3) damage or debris on the device tip between passes.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|