The patient was undergoing a thrombectomy procedure in the superficial femoral artery (sfa) using an indigo system aspiration catheter 6 (cat6) and a non-penumbra sheath.During the procedure, the cat6 became clogged, and the physician removed the cat6.Upon removal, the mid-shaft of the cat6 was observed to be ovalized.Therefore, the cat6 was not used for the remainder of the procedure.The procedure was completed using another cat6 and the same sheath.There was no report of an adverse effect to the patient.
|
Results: the cat6 was ovalized approximately 129.0, 131.0, and 132.5 thru 134.5 cm from the hub.Conclusions: evaluation of the returned cat6 confirmed that the catheter was ovalized.If the peel-able sheath (provided by penumbra) is not used to the protect the distal shaft of the catheter during insertion, and the catheter is forcefully advanced against resistance, damage such as ovalization may occur.During the functional test, the cat6 was connected to a demonstration aspiration tubing, canister, and pump.The pump was powered on and fluid was aspirated into the canister.Penumbra catheters are visually inspected during in-process inspection and during quality after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|