• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Patient Involvement (2645)
Event Date 08/10/2020
Event Type  malfunction  
Manufacturer Narrative
The primary reported complaint of the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message was confirmed during the initial functional testing and archive data review.The root cause for ua 07 was due to damaged load cells.The damaged load cells were likely due to mishandling such as drop.The secondary reported complaint of the platform displayed user advisory (ua) 45 (not at "home" position after power-on/restart) error message was not confirmed during functional testing; however, was confirmed during archive data review.The root cause for the ua45 error message was due to the driveshaft not being at home position which is likely attributed to user error.Note that the user advisory error messages are designed into the platform when one of several conditions is detected.The error message observed in the archive are easily clearable by user.For example, ua 45 alerts the operator that the autopulse driveshaft is not at its home position when the platform was powered on.This user advisory will persist until the driveshaft is returned to its home position.Per the autopulse user guide instruct, to clear ua 45, the operator needs to pull up the lifeband until the chest bands are full extended.This action will move the driveshaft to its home position.Upon visual inspection, no physical damage was observed.The archive data review showed multiple occurrences of ua07 error message and a single occurrence of ua45.The initial functional testing of the autopulse platform could not be performed due to ua 07 error message displayed upon powering on.The load cell characterization test results confirmed load cells were over reported.The load cells need to be replaced to address the ua07 error.After replacing load cells, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test battery until discharged without any fault or error.The load cell characterization test passed.The autopulse platform passed final test.Historical complaints were reviewed for service information related to the reported complaint and there was one similar complaint reported for autopulse with serial number 33933.Ccr 50355 reported on 08 july 2020, load cell 2 was replaced.
 
Event Description
During shift check, the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) and ua 45 (driveshaft not at "home" position after power-on/restart) error messages.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10492112
MDR Text Key209359735
Report Number3010617000-2020-00854
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/10/2020
Initial Date FDA Received09/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-