• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/10/2020
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of the autopulse platform (sn (b)(4)) displayed user advisory "(ua) 19" (max applied load exceeded) error message was not confirmed during the functional testing.No device malfunction was observed during the testing and the autopulse platform functions as intended.Note that the customer has tested the autopulse platform on a manikin on-site and the device functions properly without error.No physical damage was observed on the returned autopulse platform during visual inspection.However, observed the encoder drive shaft does not rotate smoothly, exhibits binding and resistance due to sticky clutch, unrelated to the reported complaint.Deburring on the clutch plate was performed to remedy this issue.The archive data could not be reviewed, because the customer deleted the archive.The returned autopulse platform passed the initial functional test without any fault or error, thus not confirming the reported complaint.After service, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test battery until discharged without any fault or error.Load cell characterization test passed.The autopulse has passed all the testing and meets all required specifications.
 
Event Description
The autopulse platform (sn (b)(4)) displayed user advisory (ua) 19 (max applied load exceeded) error message multiple times.Patient use information was requested but no additional information was provided, therefore patient use is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10492766
MDR Text Key209360648
Report Number3010617000-2020-00863
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001014
UDI-Public00849111001014
Combination Product (y/n)N
Reporter Country CodeSI
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0740-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/10/2020
Initial Date FDA Received09/04/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-