Upon initial investigations, the sample foam detection camera of the e 801 analyzer was checked and a film was clearly visible on the sample in question.Multiple samples showed film, droplets, or bubbles.As a contributing factor, the firmware of the camera had not been updated and the camera needed adjustment.
|
The initial reporter stated they received discrepant results for one patient sample processed on a cobas 8100 pre-analytics system and tested with elecsys ft3 iii and the elecsys ft4 iii assay on a cobas 8000 e 801 module.No incorrect results were reported outside of the laboratory.The sample was repeated on (b)(6) 2020 and the repeat values matched the patient's history.When the sample was first tested on the e 801 analyzer, one test parameter was flagged, indicating a liquid level detection issue.The sample was then tested again to obtain the initial ft3 and ft4 values.The initial ft3 value was 1.53 pmol/l and when repeated on (b)(6) 2020, the result was 5.64 pmol/l.The initial ft4 value of 4.37 pmol/l and when repeated on (b)(6) 2020, the result was 30.6 pmol/l.The ft3 reagent lot number was 438059, with an expiration date of 31-jan-2020.The ft4 reagent lot number was 43284402, with an expiration date of 30-sep-2020.The serial number of the e 801 analyzer is (b)(4).
|