MAUDE Adverse Event Report: GYRUS ACMI INC SHOCKPULSE LITHOTRIPSY PROBE, REUSABLE,; LITHOTRIPTOR, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI INC SHOCKPULSE LITHOTRIPSY PROBE, REUSABLE,; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-PR340
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/10/2020
Event Type  malfunction  
Manufacturer Narrative
Device was returned and evaluated at olympus malaysia service center site.Evaluation determined that the device shaft was broken.In addition, the spring was found broken and non-repairable.Both of the damages found on the device are non -repairable.Reported failure was confirmed.Likely attributed to users mishandling of the device.
 
Event Description
It was reported that the device probe was found broken due to hard and calcified stone.The issue occurred during a percutaneous nephrolithotomy (pcnl) procedure.No further details provided regarding the reported event.No patient impact or harm was reported.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the device return evaluation, complaint history review and investigation conclusion.Evaluation of the returned device confirmed the reported issue.The probe fractured approximately 6 inches from the proximal end.In addition, the spring was also broken.While a definitive root cause cannot be determined failure of this sort is often attributed to user technique.It is important the probe be held concentric to the instrument channel of the endoscope.Any applied torque may result in misalignment and contact wear leading to the fracture of the probe.As stated on the instructions for use as a preventive measure: the shockpulse-se should be placed in light contact with the stone.It is not necessary to press down on the stone with a lot of pressure.The probe is fragile.It is critical that the surgeon does not bend or torque the probes against the endoscope during the procedure.There is no need to rotate the probe or transducer; it will not improve fragmentation or stone clearance.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY PROBE, REUSABLE,
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI INC
136 turnpike road
southborough PA 01772
MDR Report Key10501620
MDR Text Key205917611
Report Number3011050570-2020-00067
Device Sequence Number1
Product Code FEO
Combination Product (y/n)N
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-PR340
Device Lot NumberD1909043-2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/11/2020
Initial Date FDA Received09/08/2020
Supplement Dates Manufacturer Received10/19/2020
Supplement Dates FDA Received10/20/2020
Is This a Reprocessed and Reused Single-Use Device? No