• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2411-36Q
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Twiddlers Syndrome (2114)
Event Date 07/09/2020
Event Type  Injury  
Manufacturer Narrative
The results/ method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Related manufacturer reference number: 2017865-2020-13184.It was reported the asymptomatic patient presented for an in clinic follow up.Upon interrogation, it was observed the atrial lead was completely dislodged with no sensing or capturing ability and implantable cardioverter defibrillator had evidence of twiddlers.The lead and device were explanted and replaced.The patient was stable.
 
Manufacturer Narrative
Analysis was normal.No device problem found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10502522
MDR Text Key205931182
Report Number2017865-2020-13183
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507615
UDI-Public05414734507615
Combination Product (y/n)N
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2015
Device Model NumberCD2411-36Q
Device Catalogue NumberCD2411-36Q
Device Lot Number4290360
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/19/2020
Initial Date FDA Received09/08/2020
Supplement Dates Manufacturer Received09/30/2020
Supplement Dates FDA Received10/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
TENDRIL STS; TENDRIL STS
Patient Outcome(s) Required Intervention;
Patient Age56 YR
-
-