It was reported that the statlock might not have enough glue and the disc allegedly disconnected from the adhesive patch of the statlock.Supposedly, the turning clamp of the statlock fell off as well.The complainant also stated that the same issue occurred twice.Per follow up via ibc on 27aug2020, the customer has confirmed that it was not the foley catheters that failed, but rather the statlock 2-way foley securing devices and the same type of device on two occasions.There was no injury to the patient, as the issue was discovered on the same day.The potential risk would be tension on the catheter if left unsecured.This could lead to bleeding, a dislodged catheter, and an urethral injury.
|
The reported event was confirmed, however the cause was unknown.Visual inspection noted that one foley statlock clamp and the swivel base was received.Visual evaluation noted that the swivel base was disconnected due to no adhesive pad being received.Although the reported event was confirmed, a root cause could not be determined.A potential root cause for this failure mode could be due to glue selection incompatible with the clamp base.The product used for the treatment purposes.The product had failed to meet the specifications, and was influenced by the reported failure.The device was not returned for evaluation.The lot number was unknown, therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "indications for use: the statlock® device is a stabilization device for compatible catheters.Contraindications: known tape or adhesive allergies.Warnings and precautions: 1.Do not use the statlock® device where loss of adherence could occur, such as with a confused patient, diaphoretic or non-adherent skin, or when the access device is not monitored daily.2.Observe universal blood and body fluid precautions and infection control procedures, during application and removal of the statlock® device.3.Minimize catheter manipulation during application and removal of the statlock® device.4.Daily maintenance: a.The statlock® device should be assessed daily and changed when clinically indicated, at least every seven days.B.If pad becomes soiled, wash with soap/water, saline or hydrogen peroxide.Do not use alcohol or prepackaged bathing systems, which could lead to early lifting.C.If showering/bathing, cover with plastic wrap or waterproof dressing.D.Conduct skin assessment prior to application and repeat daily per facility protocol.E.Use clinical judgment on the removal of the statlock® stabilization device if the patient experiences any fluid shifts that may interfere with skin integrity.Application technique prep 1.Place foley catheter into retainer.Directional arrow should point towards catheter tip, and balloon inflation arm should be next to the clamp hinge.2.Close lid, being careful to avoid pinching the catheter.3.Identify securement site by laying the device retainer on the front of the thigh, leaving 1 inch of catheter slack between insertion site and the statlock® device retainer.4.After placing the statlock® stabilization device off to the side, cleanse and degrease the securement site with alcohol per hospital policy.Let skin dry.5.Apply skin protectant, in direction of hair growth, to area larger than securement site.Allow to dry completely (10-15 seconds).6.Using permanent marker, write initials and date of application on the statlock® device anchor pad.Note: always secure catheter into the statlock® device retainer before applying adhesive pad on skin.Place and peel 7.Align the statlock® stabilization device over securement site leaving 1 inch of catheter slack.Make sure leg is fully extended.8.While holding the retainer to keep the pad in place, peel away paper backing, one side at a time and place tension-free on skin.Removal technique disengage 1.Open retainer by pressing release button with thumb, then lift to open.2.Remove foley catheter from the statlock® device.Dissolve 3.Wipe the edge of the pad using at least 5-6 alcohol pads until a corner lifts.Then continue to stroke undersurface of pad with alcohol to dissolve adhesive pad away from skin.Do not pull or force pad to remove.".
|
It was reported that the statlock might not have enough glue and the disc allegedly disconnected from the adhesive patch of the statlock.Supposedly, the turning clamp of the statlock fell off as well.The complainant also stated that the same issue occurred twice.Per follow up via ibc on 27aug2020, the customer has confirmed that it was not the foley catheters that failed, but rather the statlock 2-way foley securing devices and the same type of device on two occasions.There was no injury to the patient, as the issue was discovered on the same day.The potential risk would be tension on the catheter when left unsecured.This could lead to bleeding, a dislodged catheter, and an urethral injury.
|