The customer reported that during set up, the trima device gave a pressure test failure alarm.The customer then saw that the diversion bag was filled with air.There was not a donor connected during this event, therefore donor (patient) information is not reasonably known.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
|
Investigation: tbct japan returned the part to lakewood for investigation.The set was visually inspected for kinks, occlusions, missing parts, mis-assembly or leaks which may have contributed to the reported incident.The white sidewall pinch clamp showed some minor damage on one side.The set was received with no air in the diversion bag.All pressure sensors were inspected and determined to be functioning properly.The set was loaded onto the trima machine in the lab with clamps in their "as found" position.The set passed loading, tubing set test and made it to the connect ac stage.The procedure was halted at this point and the set was unloaded.No alarms or errors were encountered.No air went into the diversion bag during the test.Root cause: specific root cause could not be definitively determined for this incident.Possible causes include: a clamp malfunction where the clamp skews to the side as it is closed, or the clamp does not fully occlude the tubing when closed due to interference between the clamp and the tube.Additional contributing factors could be related to user interface, where either the sample bag clamp was not closed at the system prompt or the clamp was closed but it was skewed on the tubing enabling a portion of the tubing to allow air to pass.
|