• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BONE ANCHORS WITH ARTHROSCOPIC DELIVERY; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BONE ANCHORS WITH ARTHROSCOPIC DELIVERY; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 4403
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 08/14/2020
Event Type  Injury  
Event Description
It was reported that during a rotator cuff repair with regeneten, three bone anchors seemed to not line up with the punch, causing the implants to broke during insertion.All pieces were removed from the patient with a grasper.The procedure was successfully completed without significant delay using a back-up device in a different spot.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A review of the instructions for use found excessive force applied during anchor deployment may result in damage to the tissue, to the material being fixated, or to the anchor.A review of risk management files found that the reported failure was documented appropriately.Clinical evaluation found that no further medical assessment is warranted at this time.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE ANCHORS WITH ARTHROSCOPIC DELIVERY
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key10514715
MDR Text Key206329957
Report Number3003604053-2020-00099
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
PMA/PMN Number
K131635
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/21/2023
Device Catalogue Number4403
Device Lot Number2054766
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/26/2020
Initial Date FDA Received09/10/2020
Supplement Dates Manufacturer Received08/26/2020
Supplement Dates FDA Received10/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-