Device was received for evaluation.The nose cap and probe were not returned with the device.Visual inspection was performed and minor nicks, dents around the threaded part of the housing was observed.The horn tip, housing body and the o ring were found intact.The suction channel and the s knob rotation was found normal.Plug connector and switches were both observed to be normal.Transducer was tested with the reference generator and was verified that the transducer receptacle lights on generator was observed blinking to solid when transducer was plugged in.The "s" and "h" buttons of the transducer were able to activate the "standard power" and "high power" lights on the generator respectively.Each handswitch was able to latch ultrasonic output and produced low sound when double clicked "h" and "s" button.Further testing performed and the transducer passed the power test when activated with both the standard and high power modes from the reference test footswitch.Based on the evaluation findings, the reported complaint of "tip broke, threaded part damage into nose cap" was not confirmed.The transducer could be connected to the test probe and the nose cap without a problem.The device produced output and met standard specification.A root cause to the reported issue could not be conclusively determined as the nose cap or probe from the reported device was not returned.Likely probable cause of the reported issue was due to the excessive force applied on the device.
|
This supplemental report is being submitted to provide the device return evaluation conclusion, review of the device history records (dhr).The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.Device return inspection concluded that the complaint could not be confirmed.The transducer was able to be connected to the reference test probe and the nose cap without any issue.The device produced output and met standard specification.A definitive root cause could not be determined on the reported issue however, it can likely be attributed to user mishandling.The shockpulse lithotripsy probes are fragile.The probe must be kept concentric to the endoscope's instrument channel while active to prevent the introduction of unnecessary torque on the probe.Additional torque may result in damage to the device.As stated on the ifu and as a preventive measure, the user manual states: the probe is fragile.It is critical that the surgeon does not bend or torque the probes against the endoscope during the procedure.There is no need to rotate the probe or transducer; it will not improve fragmentation or stone clearance." olympus will continue to monitor complaints for this device.
|