The patient was undergoing a thrombectomy procedure in the left anterior descending (lad) artery using an indigo system catrx aspiration catheter (catrx).During the procedure, the rapid exchange port of the catrx peeled off.Therefore, the catrx and wire were removed.The procedure was completed using another catrx and wire.There was no report of an adverse effect to the patient.
|
Results: the guidewire lumen had two punctures approximately 7.0 cm from the distal tip.Conclusions: evaluation of the returned cat rx revealed catheter lumen punctures.This damage was likely the reported rapid exchange port of the cat rx peeling off.If a guidewire is inserted into the guidewire lumen at an extreme angle, damage such as punctures may occur.During functional testing, a demonstration guidewire was advanced through the cat rx guidewire lumen without an issue.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|