• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 WINDOW FRAME 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 WINDOW FRAME 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 59410882
Device Problem Loss of or Failure to Bond (1068)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/14/2020
Event Type  malfunction  
Event Description
It was reported that when the customer removed the carrier#2, the film dressing adhered to the carrier#2 and peeled off from the frame, so it was impossible to use.A backup was available.No information about delay.The sample will be returned for investigation.
 
Manufacturer Narrative
We have now concluded our investigation for the complaint received.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issues highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.A complaint history review was carried out using the part number provided, there have been no further complaints reported with this failure mode in the past three years.The device was intended to be used for treatment.As the returned dressing was disassembled, it was not possible to carry out a product evaluation.We have therefore not been able to confirm a relationship between the event and the device.Delamination is a low level intermittent fault which if seen is tabbed for removal during the conversion process.During the manufacturing of the dressing, in process checks are undertaken for this product.All material is manufactured according to the specification and only product meeting the release criteria will be passed on for further processing.We have not been able to identify a definitive root cause on this occasion.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 WINDOW FRAME 10X12CM CTN 50
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10518221
MDR Text Key206476815
Report Number8043484-2020-02820
Device Sequence Number1
Product Code KGX
UDI-Device Identifier04582111151575
UDI-Public4582111151575
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number59410882
Device Lot Number1923
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/17/2020
Initial Date FDA Received09/10/2020
Supplement Dates Manufacturer Received10/01/2020
Supplement Dates FDA Received10/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-