• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number INSTRUMENT HOLDER
Device Problem Fitting Problem (2183)
Patient Problems No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/19/2020
Event Type  Injury  
Manufacturer Narrative
Udi# : (b)(4).The device has not been evaluated yet for investigation purpose.Once the evaluation is performed, a follow-up medwatch report will be submitted.
 
Event Description
During surgery, but prior to first incision, the instrument holder (mt-02-158 s18108) could not be placed onto the end of the robot arm by the surgeon.Upon further inspection, there appeared to be notches/grooves around the center inset circle that fits onto the robot arm.The instrument holder had to be filed down around this circle inset in order to get the instrument holder to sit flush onto the robot arm.It was determined that the robot arm had no issue as the pointer probe and laser attachments sat flush with no force needed.The process of getting the instrument holder attached to the robot arm delayed surgery by 30-45 minutes while the patient was under anesthesia.
 
Manufacturer Narrative
It was reported that during surgery, the instrument holder (b)(6) could not be placed onto the end of the robot arm by the surgeon.Upon further inspection, there appeared to be notches/grooves around the center inset circle that fits onto the robot arm.Photos provided permitted to confirm the event.The most probable root cause of the damage of the instrument holder is a shock caused by a fall of the instrument or by a shock with another instrument.This cannot be confirmed.Corrected data: b4: date of this report.G3: date received by manufacturer.H2: if follow-up, what type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narratives/data.H3 other text: device was not returned to manufacturer.
 
Event Description
During surgery, but prior to first incision, the instrument holder (b)(6) could not be placed onto the end of the robot arm by the surgeon.Upon further inspection, there appeared to be notches/grooves around the center inset circle that fits onto the robot arm.The instrument holder had to be filed down around this circle inset in order to get the instrument holder to sit flush onto the robot arm.It was determined that the robot arm had no issue as the pointer probe and laser attachments sat flush with no force needed.The process of getting the instrument holder attached to the robot arm delayed surgery by 30-45 minutes while the patient was under anesthesia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
MDR Report Key10520363
MDR Text Key206493116
Report Number3009185973-2020-00207
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
PMA/PMN Number
K182417
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberINSTRUMENT HOLDER
Device Catalogue NumberROSAS00358
Device Lot NumberROSA3-275A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/19/2020
Initial Date FDA Received09/11/2020
Supplement Dates Manufacturer Received05/26/2021
Supplement Dates FDA Received05/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-