• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110137
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problems Failure of Implant (1924); No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/13/2020
Event Type  Injury  
Event Description
It was reported that during a navio procedure, there was an pfs control failure during case.The handpiece was replaced and the system was rebooted, but an error occurred during initialization: pfs control hardware failure.(errcode = 400000000000).The procedure was continued with manual instruments.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the navio handpiece, part number pfsr110137, serial (b)(6) and used for treatment, was returned for evaluation.A relationship between the reported event and the device could not be established.The reported problem could not be visually confirmed.A functional evaluation was performed.The reported problem was not confirmed.The evaluation followed a performance verification.The device performed as intended with no errors present during evaluation.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.While the reported problem was not confirmed during evaluation, the most probable contributing factor(s) to the reported problem is potential 4-10 error induced by a handpiece used prior to the one associated with this case.The surgical technique guide provides guidelines for recovering to a fully manual procedure in the ¿recovery procedure guidelines¿ per the clinical evaluation, there was no reported patient injury/impact resulted.Smith and nephew has not received adequate materials (operative reports) to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.This failure is an identified failure mode within the risk assessment released during the timeframe of the incident.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10521233
MDR Text Key206513279
Report Number3010266064-2020-01758
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2020
Initial Date Manufacturer Received 08/13/2020
Initial Date FDA Received09/11/2020
Supplement Dates Manufacturer Received11/01/2021
Supplement Dates FDA Received11/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-