• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problems Difficult to Remove (1528); Scratched Material (3020)
Patient Problem Insufficient Information (4580)
Event Date 08/10/2020
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history record could not be reviewed because the reporting facility did not provide a valid lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that during an intraocular lens (iol) implant procedure, the iol was loaded into the cartridge, the introducer passed over the lens and was unable to be inserted.It was difficult to remove the lens from cartridge and tying forceps were used to remove.The iol was reloaded and partially inserted into the patient's eye.The iol was removed again and a scratch was noted on the lens.The cartridge was reported to be faulty and the procedure was completed with a backup lens.
 
Manufacturer Narrative
There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key10521327
MDR Text Key207698276
Report Number1119421-2020-01247
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot Number15005960
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/19/2020
Initial Date FDA Received09/11/2020
Supplement Dates Manufacturer Received09/16/2020
Supplement Dates FDA Received10/06/2020
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
Treatment
DUOVISC VIS SYSTEM; MONARCH III IOL DELIVERY SYST; NORDON TYING FORCEPS; SN60WF.215, ACRYSOF SP NATURAL IQ
Patient Age82 YR
-
-