• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL AIRVO 2 HEATING BREATHING TUBE; HIGH FLOW BREATHING CIRCUIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL AIRVO 2 HEATING BREATHING TUBE; HIGH FLOW BREATHING CIRCUIT Back to Search Results
Model Number 900PT561
Device Problem Disconnection (1171)
Patient Problem Death (1802)
Event Date 07/01/2020
Event Type  Death  
Event Description
Inspiratory circuit became inadvertently disconnected during normal pt care.Disconnect was not noted until pt was in extremis.This was a high flow nasal cannula oxygen delivery system which had a device designed for the circuit which introduces neutralized medication.This device sits between the circuit and the patient nasal cannula.When the breathing circuit and the cannula are directly connected there is a locking mechanism, when the device for nebulized medication delivery is in place, there is no longer a locking mechanism.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRVO 2 HEATING BREATHING TUBE
Type of Device
HIGH FLOW BREATHING CIRCUIT
Manufacturer (Section D)
FISHER & PAYKEL
MDR Report Key10521358
MDR Text Key206706383
Report Number10521358
Device Sequence Number1
Product Code BTT
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Type of Report Initial
Report Date 07/30/2020,07/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number900PT561
Device Catalogue Number900PT561
Distributor Facility Aware Date07/04/2020
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/07/2020
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age82 YR
Patient Weight85
-
-