Model Number 74018901
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 08/17/2020
Event Type  malfunction  
Event Description
It was reported that the inner plastic of the journey tibial impl impactor has been found chipped.No case involved.
 
Manufacturer Narrative
H3, h3: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures for the listed batch; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY TIBIAL IMPL IMPACTOR
Type of Device
PROST, KNEE, PAT/FEMTIB, SEMI-CONSTRAIN, UNCEMENT, POROUS, COAT, POLY/MET/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10523552
MDR Text Key206713713
Report Number1020279-2020-04557
Device Sequence Number1
Product Code MBH
UDI-Device Identifier03596010557285
UDI-Public03596010557285
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74018901
Device Catalogue Number74018901
Device Lot Number17JSM0475
Initial Date Manufacturer Received 08/17/2020
Initial Date FDA Received09/11/2020
Supplement Dates Manufacturer Received07/30/2021
Supplement Dates FDA Received08/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-