The olympus field service engineer (fse) found the automated endoscope reprocessor's disinfectant circulation hoses needed to be replaced.There were foreign objects found in the chemical solution.The fse repair/replaced the disinfectant circulation hoses and the aer was verified according to original equipment manufacturer specifications.Additionally, software attributes have been verified and confirmed.Equipment passed the electrical safety test and the equipment functionality was verified before leaving the facility.The cause of the reported event could not be determined at this time as the investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
|