MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REFLECTION REAMER DOME 57MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REFLECTION REAMER DOME 57MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71362757
Device Problem Material Separation (1562)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/14/2020
Event Type  malfunction  
Event Description
It was reported that, during thr procedure, the "+" where the reamer dome 57 mm connects to the reamer shaft broke off.The procedure was finished using a smith and nephew backup device.Surgery was not delayed.The patient was not harmed beyond the described event.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The reamer shaft had separated from the reamer.The device was manufactured in 2019 and shows signs of extensive use.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFLECTION REAMER DOME 57MM
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10526105
MDR Text Key206712604
Report Number1020279-2020-04569
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010456953
UDI-Public03596010456953
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71362757
Device Catalogue Number71362757
Device Lot Number19CM13861
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2020
Initial Date Manufacturer Received 08/14/2020
Initial Date FDA Received09/13/2020
Supplement Dates Manufacturer Received09/03/2020
Supplement Dates FDA Received09/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-