Model Number 19AGFN-756
Device Problem Incomplete Coaptation (2507)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/04/2020
Event Type  Injury  
Manufacturer Narrative
Further information regarding this event has been requested.Investigation results will be provided in a subsequent submission.
 
Event Description
After implantation of aortic annulus, abnormal flap opening and closing was tested.
 
Manufacturer Narrative
An event of an abnormal flap opening and closing was reported.No anomalies were found with the valve leaflets, and functional testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, B.V. (CD)
lot 21
caguas west industrial park
caguas, puerto rico 00726 -099
MDR Report Key10528455
MDR Text Key206768917
Report Number3003681312-2020-00013
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005852
UDI-Public05414734005852
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Remedial Action Replace
Type of Report Initial,Followup
Report Date 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number19AGFN-756
Device Catalogue Number19AGFN-756
Device Lot Number7367083
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2020
Initial Date FDA Received09/14/2020
Supplement Dates Manufacturer Received10/12/2020
Supplement Dates FDA Received10/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-