DEPUY SYNTHES PRODUCTS LLC DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC
|
Back to Search Results |
|
Catalog Number 05.001.083 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: the device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.The manufacturing location was unknown.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the air hose device had ruptured and had component damage.It was further determined that the device failed pretest for check outer hose and general condition.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|