Model Number 1975-20-075 |
Device Problem
Device-Device Incompatibility (2919)
|
Patient Problems
Not Applicable (3189); No Code Available (3191)
|
Event Date 08/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the reclaim assembly tool for placing 2000 psi got stuck in implant after the tensile bar appropriately broke.Had to take out entire construct.Surgical delay of one hour reported.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Udi: (b)(4).Patient code: no code available (3191) is used to capture prolonged surgery and insufficient information.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
|
|