INSIGHTEC LTD EXABLATE 4000; MR GUIDED FOCUSED ULTRASOUND SYSTEM Back to Search Results
Model Number 4000
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Muscle Weakness (1967); Balance Problems (4401)
Event Date 08/18/2020
Event Type  Injury  
Manufacturer Narrative
The retrospective analysis has not indicated any technical failures or erroneous operation of the system.Treatment parameters were in line with the typical range.The system performance was found to be to spec and as expected.No new risk has been recognized.
 
Event Description
24 hours post treatment for essential tremor, patient reported leg weakness and imbalance, which was still existing 1 week after the treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXABLATE 4000
Type of Device
MR GUIDED FOCUSED ULTRASOUND SYSTEM
Manufacturer (Section D)
INSIGHTEC LTD
5 nachum heth street
tirat carmel, 39120
IS  39120
Manufacturer (Section G)
INSIGHTEC LTD
5 nachum heth street
tirat carmel, 39120
IS   39120
Manufacturer Contact
meital trank
5 nachum heth street
tirat carmel, 39120
IS   39120
MDR Report Key10539076
MDR Text Key207104388
Report Number9615058-2020-00018
Device Sequence Number1
Product Code POH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4000
Device Catalogue NumberSYS940000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/19/2020
Initial Date FDA Received09/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Disability;
-
-