• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems High impedance (1291); Therapy Delivered to Incorrect Body Area (1508)
Patient Problems Undesired Nerve Stimulation (1980); Complaint, Ill-Defined (2331)
Event Date 09/15/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 4351-35, serial#: (b)(4), implanted: (b)(6) 2019, product type: lead.Product id: 4351-35, serial#: (b)(4), implanted: (b)(6) 2019, product type: lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 29-jul-2021, udi#: (b)(4), product id: 4351-35, serial/lot #: (b)(4), ubd: 29-jul-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a health care provider (hcp) regarding a patient who was implanted with a neurostimulator for gastric stimulation.  it was reported the impedance issues on electrode 2 >2 ,000 ohms.Caller programmed c+ and 3- but patient reported stimulation in abdomen.Reviewed possible flipped ins which can create intra-muscular stimulation.Also discussed that may be reason for lead impedance.Suggested x-ray for ins position and reducing amplitude to reduce,or alleviate stimulation sensation.Caller reported a year ago, impedance was 1,044 ohms.Patient has been having decreased symptoms in the past year, but then reported patient seems to have difficultly in general, and mentioned if the patient had an endoscopy they would probably see lead migration.No further complications were reported/anticipated at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key10541590
MDR Text Key207182820
Report Number3004209178-2020-16075
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2020
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/15/2020
Initial Date FDA Received09/16/2020
Date Device Manufactured05/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
-
-