Catalog Number 13546-59S
Device Problem Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Date 08/20/2020
Event Type  malfunction  
Event Description
It was reported that the middle port in front on the left side of the controller (foot port) was damaged.No case involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10: h3, h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A visual inspection observed the footswitch port is pushed inside the unit and the lid and bezel are damaged.A functional evaluation revealed most settings cannot be tested due to the footpedal port being pushed inside the unit.The unit was opened and found the footpedal port and its hardware loose inside the unit.The complaint was verified.Factors, unrelated to the design or manufacture of the device which could have contributed to the complaint event, include misalignment of the connector when connecting to the unit receptacle in which excessive force or twisting could cause damage.No containment or corrective actions are recommended at this time.There are no indications to suggest that the device/product did not meet specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CII (120V) SRVC, CONTROLLER-ONLY PACKAGE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10544804
MDR Text Key207299938
Report Number3006524618-2020-00781
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number13546-59S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2020
Initial Date Manufacturer Received 08/24/2020
Initial Date FDA Received09/17/2020
Supplement Dates Manufacturer Received10/05/2020
Supplement Dates FDA Received10/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-