• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNK FIXATION PIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNK FIXATION PIN Back to Search Results
Catalog Number UNK FIXATION PIN
Device Problems Device-Device Incompatibility (2919); Mechanical Jam (2983)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 09/03/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the pin got stuck in a pin hole during the procedure, and could not be removed.Had to cut both ends of the pin off, and leave the rest of it stuck in the block in order to proceed.No surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirms the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK FIXATION PIN
Type of Device
FIXATION PIN
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10546850
MDR Text Key208217096
Report Number1818910-2020-20331
Device Sequence Number1
Product Code HTY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK FIXATION PIN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2020
Initial Date Manufacturer Received 09/03/2020
Initial Date FDA Received09/17/2020
Supplement Dates Manufacturer Received09/12/2020
09/23/2020
Supplement Dates FDA Received09/21/2020
09/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age61 YR
-
-