• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TANDEM BIPOLAR COCR 46OD 28ID; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TANDEM BIPOLAR COCR 46OD 28ID; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED Back to Search Results
Model Number 71322046
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Injury (2348)
Event Date 09/27/2018
Event Type  Injury  
Event Description
It was reported that revision was performed due to painful right hip bipolar arthroplasty with erosive changes of the acetabulum.Taper femoral head, shell/ uhmwpe liner and femoral component were exchanged and procedure was converted to a total hip replacement.Stem was retained.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, the root cause of the ¿erosion¿ of the acetabular region cannot be definitively concluded and it cannot be ruled out that progression of disease played a role and therefore not a failure of a smith and nephew device.Without x-rays or the explanted components a full investigation cannot be made.The activity of the patient is unknown and the prosthesis was insitu about eleven years.A review of the complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness or wear.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TANDEM BIPOLAR COCR 46OD 28ID
Type of Device
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10551840
MDR Text Key207492885
Report Number1020279-2020-04658
Device Sequence Number1
Product Code KWY
UDI-Device Identifier03596010492456
UDI-Public03596010492456
Combination Product (y/n)N
PMA/PMN Number
K023743
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71322046
Device Catalogue Number71322046
Device Lot Number07BM07767
Initial Date Manufacturer Received 08/20/2020
Initial Date FDA Received09/18/2020
Supplement Dates Manufacturer Received08/25/2021
Supplement Dates FDA Received08/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-