Catalog Number 04625374160
Device Problem Non Reproducible Results (4029)
Patient Problem Coagulation Disorder (1779)
Event Date 08/21/2020
Event Type  malfunction  
Manufacturer Narrative
The reporter's meter and test strips were provided for investigation where they were tested using retention controls.Testing results (qc range = 2.3 - 3.5 inr): qc 1: 2.8 inr, qc 2: 2.7 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." routine retention testing is performed.Retention testing data is reviewed, and appropriate actions are taken as needed.Occupation was lay user/patient.
 
Event Description
The initial reporter received questionable results from coaguchek xs meter serial number (b)(4).At 2 pm, the laboratory result using an unknown reagent was 2.1 inr.At 4:40 pm, the meter result was 1.2 inr.The laboratory result was believed to be correct and no treatment was received.The therapeutic range was 2.0-3.0 inr and the customer tests every 3 weeks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10553394
MDR Text Key207736835
Report Number1823260-2020-02297
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2021
Device Catalogue Number04625374160
Device Lot Number43002022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2020
Initial Date Manufacturer Received 08/21/2020
Initial Date FDA Received09/18/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age77 YR
Patient Weight71
-
-