Impacted device: sd08r-sp05x-000 lot#726727728 wo#135028.The preliminary risk assessment was conducted for this issue on (b)(6) 2020.According to the assessment and as documented in the event description, the customer stated they decided not to use the electrode with the bent end before the first contact as they did not want to risk the electrode deviating from the target location due to the bend.Because of the concern for a trajectory deviation through potentially healthy brain tissue if the electrode were to be used, the potential harm of damage to healthy tissue remains.Therefore, mechanical energy, with a severity of 4 was used.Additionally, in sept 2020 a medical expert was contacted via email to determine the likelihood of a bent tip impacting patient safety.The medical expert confirmed that this type of bend could cause injury.
|