Model Number 5MAXJET7KIT-B |
Device Problems
Break (1069); Suction Problem (2170); Physical Resistance/Sticking (4012)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/27/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
The patient was undergoing a thrombectomy procedure in the in the middle cerebral artery (mca) using a penumbra system jet 7 reperfusion catheter (jet7) and a neuron max 6f 088 long sheath (neuron max).During the procedure, the physician completed one pass using the jet7 and neuron max.While advancing the jet7 for a second pass, the physician experienced resistance and no aspiration was observed during the second pass.Therefore, the jet7 was removed.After the removal, it was observed that the distal end of the jet7 was fractured.The procedure was completed using another jet7 and the same neuron max.There was no report of an adverse effect to the patient.
|
|
Manufacturer Narrative
|
Please note that the device associated with this complaint was expected to be returned; however, additional information received from the penumbra sales representative indicated that the device was disposed of and is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|
|
|