Investigation completed on a smiths medical syringe infusion pumps|medfusion 4000 pumps.The complaint of force sensor error was confirmed in the event log and during testing to duplicate event this revealed force sensor values were out of spec; values were 0= -1.28 lb, 5= 3.13 lb, 15= 12.71 lb.Physical condition of the device revealed, cracked battery door and right plunger case.The event if believed to be caused from force sensor was out of calibration due to normal wear.No visible damage found on force sensor.Action was taken to replace the force sensor and calibrate the device.Other maintenance included: replaced right plunger case due to physical damage, replaced bottom case due to broken seal.Replaced worm gear and motor mount due to preventative action.Replaced cable guard due to missing.Cleaned, greased; calibrated device and performed all functional tests which passed.
|