SMITH & NEPHEW, INC. FEMORAL IMPLANT IMPACTOR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71440005
Device Problem Material Separation (1562)
Patient Problem No Patient Involvement (2645)
Event Date 08/24/2020
Event Type  malfunction  
Event Description
It was reported that the rubber part of the impactor came off of it.This was found during set up / inspection.No surgery or patient involvement.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The plastic on the cam arms are missing from the device, rendering the device inoperable.The device was manufactured in 2008 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL IMPLANT IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10558075
MDR Text Key207696280
Report Number1020279-2020-04710
Device Sequence Number1
Product Code JDG
UDI-Device Identifier00885556034156
UDI-Public00885556034156
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440005
Device Catalogue Number71440005
Device Lot Number08EM15617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2020
Initial Date Manufacturer Received 08/24/2020
Initial Date FDA Received09/20/2020
Supplement Dates Manufacturer Received11/13/2020
Supplement Dates FDA Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-