BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number 8900
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/23/2020
Event Type  malfunction  
Manufacturer Narrative
Device returned to manufacturer: the tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The device showed three kinks located 27.5cm, 111.5cm and 161cm from the tip.The tip showed bend damage.There was peeled coating at the 111.5cm location.The occ handle was connected to the ffr link to verify the signal strength.There were no issues in connecting to the ffr link.The signal was present and showed green lights as designed.The occ handle was then connected to the bench top testing equipment and the wire was inserted into the pressure chamber.The pressure was increased to verify the sensor was indeed reacting to the pressure increases and decreases.The pressure sensor functioned as designed.The coefficient values were confirmed to be programmed per document.The occ handle was again connected to the ffr link.The device was then connected to the polaris (ilab) test equipment via bluetooth signal.The wire communicated to the polaris system and zeroed as designed.With the wire inserted into the test pressure chamber, the wire transferred a pressure waveform to the polaris which indicates a functioning wire.The wire was removed from the occ handle with no issues.The sensor port showed residue of body fluids.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The complaint was not confirmed for drift issues.
 
Event Description
Reportable based on analysis completed 4 sept 2020.It was reported that a drift in pressure value occurred with the comet pressure guidewire.There were no patient complications reported.However, upon receiving the device, peeling of the coating was noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10559913
MDR Text Key207718861
Report Number2134265-2020-13032
Device Sequence Number1
Product Code DXO
UDI-Device Identifier08714729904403
UDI-Public08714729904403
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/25/2022
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0025405179
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/04/2020
Initial Date FDA Received09/21/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-