NIHON KOHDEN CORPORATION GZ-130PA; TRANSMITTER Back to Search Results
Model Number GZ-130PA
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/27/2020
Event Type  malfunction  
Manufacturer Narrative
The customer reported that their vital sign telemetry is experiencing continuous dropouts from the network.No harm or injury was reported.The customer will be sending this particular unit in for an exchange.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.The following fields contains no information (ni), as attempts to obtain information were made, but not provided.Patient information.
 
Event Description
The customer reported that their vital sign telemetry is experiencing continuous dropouts from the network.
 
Manufacturer Narrative
Details of complaint: the customer reported that the vital sign telemetry was experiencing continuous dropouts from the network.No patient harm or injury was reported.Investigation summary: nihon kohden technical support (nk ts) initiated an exchange.With the available information, a root cause cannot be determined.Device received into nk warehouse.The exchange unit was installed, and the issue was resolved.Comm loss is an error message displayed on individual bed tiles on the cns that alerts clinicians that the connection between the cns and the device it is monitoring has been lost.Potential causes of comm loss: network issues: the access points in the area where the issue is occurring may have been loaded with traffic or have a bad connection, or they may have experienced interference in the area causing the connectivity issues.If the network host table becomes unbalanced, comm loss could occur.If the telemetry device is approaching the range boundaries or out of range of an access point, it may experience comm loss.
 
Event Description
The customer reported that the vital sign telemetry was experiencing continuous dropouts from the network.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GZ-130PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 116-8 560
JA  116-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key10574364
MDR Text Key208159270
Report Number8030229-2020-00662
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921117415
UDI-Public04931921117415
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153707
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGZ-130PA
Device Catalogue NumberGZ-130PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/27/2020
Initial Date FDA Received09/23/2020
Supplement Dates Manufacturer Received02/23/2023
Supplement Dates FDA Received03/02/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/29/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
NA.
-
-