RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Energy Output Problem (1431)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 23sep2020.
 
Event Description
It was reported to philips that the device had an over voltage protection issue.The device was not being used on a patient at the time of the event.There was no patient or user harm reported.
 
Manufacturer Narrative
G4: 06nov2020 b4: (b)(6) 2020 an authorized service personnel(asp) was dispatched to the customer site and it was determined that the power management printed circuit board assembly (pm pcba) needed to be replaced to return the device to working condition.The asp replaced the pm pcba.The device passed performance verification testing.Following the repair, the device was returned to the customer to be placed back into service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10577929
MDR Text Key208724281
Report Number2031642-2020-03433
Device Sequence Number1
Product Code MNT
UDI-Device Identifier884838025776
UDI-Public(01)884838025776
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 09/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2020
Initial Date FDA Received09/23/2020
Supplement Dates Manufacturer Received09/05/2020
Supplement Dates FDA Received11/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-