Model Number 71360093
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 08/31/2020
Event Type  malfunction  
Event Description
It was reported that the head of the impactor was broken and need to be a replacement.Backup from smith & nephew was available.No delay.Not any patient involved in the case.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H6: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD/NECK IMPACTOR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10583879
MDR Text Key208486821
Report Number1020279-2020-04852
Device Sequence Number1
Product Code LZO
UDI-Device Identifier03596010619310
UDI-Public03596010619310
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360093
Device Catalogue Number71360093
Device Lot Number15BM15721
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/31/2020
Initial Date FDA Received09/24/2020
Supplement Dates Manufacturer Received04/27/2021
Supplement Dates FDA Received04/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-