SMITH & NEPHEW, INC. DRILL 3.2MM BLUE; TEMPLATE Back to Search Results
Model Number 00579107100
Device Problem Dull, Blunt (2407)
Patient Problem No Patient Involvement (2645)
Event Date 09/01/2020
Event Type  malfunction  
Event Description
It was reported that a drill 3.2mm blue was dull.No case involved, therefore there was no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The cutting edges of the drill are rendering the device inoperable.The device was manufactured in 2016 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL 3.2MM BLUE
Type of Device
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10584096
MDR Text Key208487087
Report Number1020279-2020-04853
Device Sequence Number1
Product Code HWT
UDI-Device Identifier00885556619056
UDI-Public00885556619056
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00579107100
Device Catalogue Number00579107100
Device Lot Number63311903
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Initial Date Manufacturer Received 09/01/2020
Initial Date FDA Received09/24/2020
Supplement Dates Manufacturer Received01/11/2021
Supplement Dates FDA Received01/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-