Catalog Number A066100000300 |
Device Problem
Break (1069)
|
Patient Problems
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
While no serious injury resulted in this event, there has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it was reported that a start-x tip broke during use; no injury resulted the broken parts have not been retrieved from the patient's mouth.
|
|
Manufacturer Narrative
|
Returned start-x-tip ems insert 3 is actually broken in the active part.No material defect was found during analysis of the rupture pattern.No unused device is available for possible evaluation.The batch number is unknown, dhr cannot be reviewed.No clear information was given regarding technique (percentage instead of digits), we cannot rule on its compliance with maillefer's recommendations.Root causes are not identified.We will track this kind of event and monitor the trend.
|
|