Model Number UHI-4 |
Device Problems
Defective Device (2588); Device Sensing Problem (2917)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/27/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not returned to olympus medical systems corp.(omsc) for evaluation.There were no further details provided.If significant additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that cylinder pressure indicator failure of the front panel of the device occurred.Other detailed information was not provided.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer¿s final investigation.The device was returned to an olympus service center for evaluation and it was confirmed that the pressure indicator was defective.This was attributed to normal wear due to the age of the device.Based on the legal manufacturer¿s investigation, the device history record (dhr) was reviewed and there were no problems found during the manufacturing of the device.The device met all specifications at the time of shipment.
|
|