MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DISCOVERY ELBOW; DISC ULNA 2.5X84MM LT BOND CT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DISCOVERY ELBOW; DISC ULNA 2.5X84MM LT BOND CT Back to Search Results
Catalog Number 114806
Device Problems Break (1069); Appropriate Term/Code Not Available (3191)
Patient Problem Unspecified Infection (1930)
Event Date 08/28/2020
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery - infection and broken ulna bone.
 
Manufacturer Narrative
The reason for this revision surgery was reported as infection and broken ulna bone.The previous surgery and the surgery detailed in this event occurred 46 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection and broken ulna bone.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection (inhibited the patient's immune system) or the broken bone.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC ULNA 2.5X84MM LT BOND CT
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key10586499
MDR Text Key208558460
Report Number1644408-2020-00837
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
PMA/PMN Number
K090473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/04/2021
Device Catalogue Number114806
Device Lot Number921520
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/28/2020
Initial Date FDA Received09/25/2020
Supplement Dates Manufacturer Received10/23/2020
Supplement Dates FDA Received11/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
114908 LOT 548180; 540-00-000 LOT 84174; 114908 LOT 548180; 540-00-000 LOT 84174
Patient Outcome(s) Required Intervention;
Patient Age54 YR
-
-