MEDOS INTERNATIONAL SàRL CH UNKNOWN RODS; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Device Problem Migration or Expulsion of Device (1395)
Patient Problems Failure of Implant (1924); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
510k: this report is for an unknown rod/unknown lot.Part and lot number are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date the surgeon was going to revise a previous construct where he used the expedium verse.One rod of this two rod construct had migrated superiorly.Upon exposure it was determined that all of the verse set screws were not tight, not just the side where the rod had migrated, but all the unitized set screws on both rods.Titanium 5.5 rods were used in the original surgery and all unitized set screws were tightened using the dss supplied torque limiting driver.The procedure and patient outcome are unknown.This report is for one (1) unknown rod.This is report 3 of 4 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Concomitant device: unknown torque device (part unknown, lot unknown, quantity unknown).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN RODS
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10586730
MDR Text Key208550339
Report Number1526439-2020-01797
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/26/2020
Initial Date FDA Received09/25/2020
Supplement Dates Manufacturer Received10/12/2020
Supplement Dates FDA Received10/13/2020
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-