• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING (PS) LEFT 13 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING (PS) LEFT 13 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Tissue Damage (2104)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2020-00238, 0001822565-2020-03346, and 0002648920-2020-00430.Concomitant medical devices: femur cemented posterior stabilized (ps) narrow left size 8 catalog#: 42500006401 lot#: 64558631, natural tibia trabecular metal two-peg porous fixed bearing left size d catalog#: 42530006701 lot#: 64678941, all poly patella cemented 32 mm diameter catalog#: 42540000032 lot#: 64620936.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that patient developed an allergic reaction and blisters on the surgical site subsequent to bilateral total knee arthroplasty.Topical medications were prescribed.No revision has been reported.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of medical records.Review of the available records identified the following: three weeks after implantation: allergic reaction at knee resulting in rash and blisters, topical otc benedryl and hydrocortisone prescribed, outcome pending.Six week visit ¿ normal alignment and stability.Rom 100.6-10 degrees flexion contracture.Moderate pain.Normal radiographic findings.Two month office visit ¿ overall, patient reports her pain is improving, has less stiffness, and is pleased with her progress.Ambulating with appropriate gait with cane and walker.The incision is well healed and intact.Rom 3-115.The knee is stable with no laxity noted with valgus or varus stressing.Quadriceps strength 5/5.Neurovascularly intact.No x-rays on the visit.Previous x-rays have shown a well-positioned, stable arthroplasty.Surgeon¿s recommendations ¿ d/c formal therapy.Transition from cane/walker to cane only.Released to drive.F/u one month with plan to return to work.Device history record was reviewed and no discrepancies were found.The allergic reaction was most likely caused by the patient's allergies to wellbutrin, latex, and pcn since it was treated with otc meds; however it cannot be confirmed with the information available.Therefore, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that patient developed an allergic reaction and blisters on the surgical site subsequent to bilateral total knee arthroplasty.Topical medications were prescribed.Outcome has since resolved.Attempts to obtain additional information have been made; however, no more is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING (PS) LEFT 13 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key10592511
MDR Text Key208721342
Report Number3007963827-2020-00239
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
PMA/PMN Number
K113369
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Type of Report Initial,Followup
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42511400513
Device Lot Number64444409
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/08/2020
Initial Date FDA Received09/28/2020
Supplement Dates Manufacturer Received12/10/2020
Supplement Dates FDA Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10
Patient Outcome(s) Required Intervention;
Patient Weight80
-
-