Model Number V60
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 28sep2020.
 
Event Description
It was reported to philips that the device had a would not operate with the battery.The device was not being used on a patient at the time of the event.There was no patient or user harm reported.
 
Manufacturer Narrative
G4: 04jan2021 b4: (b)(6) 2021.The customer evaluated the device with assistance from a philips remote service engineer (rse), and it was determined that the battery needed to be replaced to return the device to working condition.The customer¿s bio-med replaced the battery to resolve the issue and bring the device back to functionality.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key10597188
MDR Text Key209573245
Report Number2031642-2020-03468
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/09/2020
Initial Date FDA Received09/28/2020
Supplement Dates Manufacturer Received09/09/2020
Supplement Dates FDA Received01/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-