• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY TRACER METRO DIRECT WIRE GUIDE; OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY TRACER METRO DIRECT WIRE GUIDE; OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number METII-35-480
Device Problem Positioning Failure (1158)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/25/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter occupation: unknown.Investigation evaluation: our evaluation of the product said to be involved confirmed the report.The wire guide is bent at approximately 4.5c m from the distal end.Rough spots on the coating were also found throughout the length of the wire guide.When functionally tested with a compatible cook sphincterotome, slight resistance was felt while advancing wire guide.No other anomalies were detected with the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use instruct the user to do the following: "for best results, wire guide should be kept wet." failure to flush the wire guide can result in damage to the wire guide.Prior to distribution, all tracer metro direct wire guides are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook tracer metro direct wire guide.The physician advanced the device to the desired position and felt obvious resistance while exchanging the sphincterotome through the wire guide.The physician retracted the sphincterotome and the wire guide from the patient, and changed to another one of the same wire guide to complete the procedure.This event was not reportable at this time.Clarifying information received 31-aug-2020 that the wire guide and sphincterotome were both removed because the wire guide became stuck and was unable to move [loss of wire guide access, subject of the report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACER METRO DIRECT WIRE GUIDE
Type of Device
OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key10597334
MDR Text Key218728451
Report Number1037905-2020-00412
Device Sequence Number1
Product Code OCY
UDI-Device Identifier10827002256863
UDI-Public(01)10827002256863(17)230414(10)W4340141
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2023
Device Catalogue NumberMETII-35-480
Device Lot NumberW4340141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2020
Initial Date Manufacturer Received 08/31/2020
Initial Date FDA Received09/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient Weight58
-
-