• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Model Number IPN038182
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/05/2020
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
Issue reported: defective lor syringe loss of resistance.
 
Event Description
Issue reported: defective lor syringe loss of resistance.
 
Manufacturer Narrative
Qn#(b)(4).A device history record review was performed on the lor syringe with no relevant findings.A design history review was performed for part # kz-05501-002 as a part of this complaint investigation.Per (b)(4) (released 12-mar-2020), supplier (preox) made the following changes: optional component materials/mold locations: option 1: barrel: polypropylene - profax pf-535 lyondell-basell.Plunger: polypropylene 100-ga12 ineos olefins & polymers (molded at fleima plastic).Blue stopper: silicone rubber (molded at et elastomer technik).Option 2: barrel: polypropylene - profax pf-535 lyondell-basell.Plunger: polypropylene profax pf-531 lyondell-basell (molded at gpe).Blue stopper: silicone rubber (molded at psilkon).Lubricant: - the i.D.Of the barrel lubricated w/ medical grade silicone (polydimethylsiloxan) and amount will not exceed 0.25mg per sq centimeter.These effective changes did impact product design and material.A corrective action is not required at this time as the root cause for this complaint investigation could not be determined based upon the information provided and without the actual sample.Complaint verification testing could not be performed as no sample was provided for analysis.A device history record review was performed on the lor syringe with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without the sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Type of Device
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key10599736
MDR Text Key208964804
Report Number3006425876-2020-00836
Device Sequence Number1
Product Code BSO
Combination Product (y/n)N
PMA/PMN Number
K140110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/10/2022
Device Model NumberIPN038182
Device Catalogue NumberJC-05400-E
Device Lot Number71F20E1066
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/10/2020
Initial Date FDA Received09/29/2020
Supplement Dates Manufacturer Received10/28/2020
Supplement Dates FDA Received11/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-