Catalog Number 204H0300
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
Type of the device: product code: qhp.Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that a patient will need a revision surgery after the their curve began to increase 3 months postoperatively.A revision is planned, to remove and retighten the cord, but not scheduled at this time.No further patient or surgical information has been provided.
 
Event Description
It was reported that a patient will need a revision surgery after the their curve began to increase 3 months postoperatively.A revision is planned, to remove and retighten the cord, but not scheduled at this time.No further patient or surgical information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A2: date of birth: unknown day in (b)(6) of 2007 a4: weight: 33.8 kg d4: udi#: (b)(4).D11: catalog#: 211h6035 6.0mm dia x 35mm vbt assy, lot#: ni.Catalog#: 211h6032 6.0mm dia x 32.5mm vbt assy, lot#: 2958624.Catalog#: 302h0012, lot#: 2958624.Catalog#: 211h6030 6.0mm dia x 30mm vbt assy, lot#: 2958623.Catalog#: 211h6025 6.0mm dia x 25mm vbt assy, lot#: 2970357.Catalog#: 211h6027 6.0mm dia x 27.5mm vbt assy, lot#: 2958622.The event is confirmed with x-rays received.Visual examination of the x-rays confirmed that the curve of the patient's spine had exceeded 38 degrees post-op.Dhr was reviewed and no discrepancies relevant to the reported event were found.Following the initial surgery, patient condition should have stabilized or improved.Although there is no clear defect with the part, the patient's condition did not appear to improve.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORD 300MM
Type of Device
THE TETHER- VERTEBRAL BODY TETHERING SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
MDR Report Key10600291
MDR Text Key209003199
Report Number3012447612-2020-00574
Device Sequence Number1
Product Code QHP
Combination Product (y/n)N
PMA/PMN Number
H190005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Catalogue Number204H0300
Device Lot Number2966796
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/31/2020
Initial Date FDA Received09/29/2020
Supplement Dates Manufacturer Received09/16/2020
09/16/2020
Supplement Dates FDA Received10/14/2020
10/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age12 YR
Patient Weight33
-
-