FISHER & PAYKEL HEALTHCARE LIMITED VENTED AUTOFEED HUMIDIFICATION CHAMBER; BTT Back to Search Results
Model Number MR290V
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was not returned to fisher & paykel healthcare for evaluation.Our investigation is thus based on the information and videographs provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the provided videograph revealed that the mr290v chamber dome was leaking water between chamber dome and base.Conclusion: we are unable to determine what caused the reported event.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "set appropriate ventilator alarms." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.".
 
Event Description
A distributor reported on behalf of a healthcare facility in (b)(6) via a fisher & paykel healthcare (f&p) field representative that a mr290v vented autofeed humidification chamber was leaking water prior to use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key10602147
MDR Text Key212142886
Report Number9611451-2020-00834
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial
Report Date 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number200203
Initial Date Manufacturer Received 08/31/2020
Initial Date FDA Received09/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-