• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL STEM 12/14 NECK TAPER PLASMA SPRAYED; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL STEM 12/14 NECK TAPER PLASMA SPRAYED; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Material Erosion (1214)
Patient Problems Erosion (1750); Pain (1994); Local Reaction (2035)
Event Date 05/13/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical devices: cat# 00875705801 shell with cluster holes porous lot# unk; cat# 00877001340 metasul taper liner ll/40 lot#unk; cat# 00877004003 metasul head 40, 12/14, size l /+3.5 lot#unk.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2020 - 03374.
 
Event Description
It was reported patient¿s legal counsel reported patient underwent right hip revision approximately 5 years post implantation due to patient allegations of pain, elevated metal ion levels.Operative records received show altr and bone erosions around the well ¿ fixed femoral and acetabular components.The head and liner were revised.No additional information.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of medical records.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: the patient underwent an initial right total hip arthroplasty due to degenerative joint disease.Zimmer biomet components were implanted without any complications.The patient was revised due to failed hip arthroplasty.During the procedure, yellow turbid non purulent fluid was encountered.Adverse tissue reaction material had eroded along the very proximal aspect of the bone-prosthetic interface of the proximal femoral stem.The acetabular cup was well fixed with small erosions of adverse tissue reaction around the margin.The head and liner were replaced.No other findings/complications related to the reported event were noted.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM 12/14 NECK TAPER PLASMA SPRAYED
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10605272
MDR Text Key209129197
Report Number0001822565-2020-03379
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
PMA/PMN Number
K161830
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Model NumberN/A
Device Catalogue Number00771100900
Device Lot Number61356526
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/25/2020
Initial Date FDA Received09/30/2020
Supplement Dates Manufacturer Received10/01/2020
Supplement Dates FDA Received10/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient Outcome(s) Hospitalization; Required Intervention;
-
-